2. Carefully detach the Force Display Module (front half) from the Force
Cell Module (rear half) and set it aside.
3. Select the Force Cell Module you intend to use.
4. Carefully align the center connectors and press the modules together.
5. Replace the four screws. Proceed to the next Section - G.3.
CAUTION: When "HELP" is displayed it may indicate either:
1) Force Cell Module is not securely connected.
2) Force Cell Module has been overloaded and is damaged.
Call Wagner Instruments for instructions.
G.3 Selecting the Load Cell Capacity
1. Enter the Set-Up Menu
a) Turn off the FDV
b) Press and hold SELECT
c) Press POWER
d) Release SELECT and POWER
NOTE: An "Access to Set-up Menu" label is provided on the back of the
FDV gage for quick reference.
2. Scroll through the menu to the Load Cell Option (LoAd) and press
the SELECT button.
3. Use the SCROLL button to step through the load cell capacities:
11pound capacity (not presently available)
22pound capacity (not presently available)
55pound capacity
10 10 pound capacity
30 30 pound capacity
50 50 pound capacity
100 100 pound capacity