15
period of inactivity (a period during which no buttons are
pressed). The period of time can be selected as follows:
no FDV does not shut off automatically
1 shuts off after 1 minute of inactivity
5 shuts off after 5 minutes of inactivity
10 shuts off after 10 minutes of inactivity
20 shuts off after 20 minutes of inactivity
30 shuts off after 30 minutes of inactivity
4. Digital Filter (Fltd)
Minor vibrations and electrical background noise in the test area, not
related to the test, may be detected by the FDV. If it is impossible to
isolate the FDV from these vibrations and/or electrical noises the
FDV readings may be erratic. The FDV features four (4) levels of
filtering for both the displayed and peak readings to minimize the
effects of background noise.
a) Filtering for Current Readings - The level of filtering required for
currently displayed readings is quite similar to that required for peak
readings. In this case however, the filtering for currently displayed
readings will affect how frequently the display will be updated.
Filtering of currently displayed readings may be set as follows.
Fc1 Average of 1 sample (fast display update)
Fc2 Average of 2 samples
Fc4 Average of 4 samples
Fc8 Average of 8 samples (slow display update)
b) Filtering for Peak Readings - The level of filtering required for peak
readings is dependent on the nature of the test and the surrounding
environment. Filtering will affect the maximum peaks recorded. To
capture critical peak force, the lowest level of filtering (FP1) is
recommended. If the gage is used in an environment with excessive
background noise, the highest level of filtering (FP8) is recommended.